Products

Specify Selection

digital DVGW compliant UV Sensor with RS485 (ModBus/ZCON)

Features

  • sensors are intended for use in certified systems in compliance to
    - German standard DVGW W294
    - Austrian standards ÖNORM M5873-1 and Vornorm ÖNORM M5873-2 The D-SiCONORM sensors are certified by an ÖNORM accredited laboratory.
  • Note: According to DVGW/ÖNORM these UV-sensors are to be used with a measurement window (e.g. ZED MF001) to allow the demanded calibration checks; ZED offers a sensor recalibration service.
  • detecting absolute UV irradiance values
  • UV values are indicated in "W/m²", "mW/cm²" or "%" with the appropriate UV-monitor
  • available for low pressure and medium pressure applications
  • all digital UV-sensors are adjusted to cover the complete specified measurement range
    - low pressure types: range 2...500W/m²
    - medium pressure types: range 20...3000W/m²
    - customized ranges are available on request
  • connection of additional sensors to the RS485 bus allows multi sensor operation

Technical Specification

supply voltage: 12...24V DC
sensitive element: silicon carbide diode (SiC)
spectral range: 210 - 290nm
operation temperatur: 0 - 60°C (32 - 140°F)
max. pressure / IP code: not waterproof without measurement window MF001 IP64 on terminal ends if connected to appropriate sensor cable
body material: stainless steel 1.4305
mounting: use with MF001 (MF001: pipe thread ISO228 G1)
wiring: ready-to-use ZED sensor cables available in several lengths

Display Units & Accessories



Products


We use cookies

We use cookies on our website. Some of them are essential for the operation of the site, while others help us to improve this site and the user experience (tracking cookies). You can decide for yourself whether you want to allow cookies or not. Please note that if you reject them, you may not be able to use all the functionalities of the site.